Semiautomated external defibrillator errors in recognizing ventricular fibrillation

Ballesteros Peña S, Lorrio Palomino S, Rollán Vallejos J


Affiliation of the authors

Bilbao SAMUR-Protección Civil, Ayuntamiento de Bilbao, Spain. Emergenciak-Osakidetza, Spain.

DOI

Quote

Ballesteros Peña S, Lorrio Palomino S, Rollán Vallejos J. Semiautomated external defibrillator errors in recognizing ventricular fibrillation. Emergencias. 2013;25:119-22

Summary

We report 3 cases of semiautomated external defibrillator error during analysis of

ventricular fibrillation. The first involved operator-dependent error, in which ambulance

motion led to failure to order a shock. The other 2 cases involved defibrillator-dependent

error, in which signals from a pacemaker apparently led to the automatic decision not to

shock even when ventricular fibrillation was detected.

 

More articles by the authors

Leave a Reply

Your email address will not be published. Required fields are marked *

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.