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
Summary
Semiautomated external defibrillator errors in recognizing ventricular fibrillation
Ballesteros Peña S, Lorrio Palomino S, Rollán Vallejos J
Affiliation of the authors
Bilbao SAMUR-Protección Civil, Ayuntamiento de Bilbao, Spain. Emergenciak-Osakidetza, Spain.
DOI
Quote
Ballesteros Peña S, Lorrio Palomino S, Rollán Vallejos J. Semiautomated external defibrillator errors in recognizing ventricular fibrillation. Emergencias. 2013;25:119-22
Summary
We report 3 cases of semiautomated external defibrillator error during analysis of
ventricular fibrillation. The first involved operator-dependent error, in which ambulance
motion led to failure to order a shock. The other 2 cases involved defibrillator-dependent
error, in which signals from a pacemaker apparently led to the automatic decision not to
shock even when ventricular fibrillation was detected.